Ametek Lx Series II Programming Manual Manual do Utilizador Página 216

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 242
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 215
Programming Manual Lx \ Ls Series II
212
8.7.4 MS704 Abnormal State Commands
Abnormal state tests can be started by sending one of the MS704:ABN commands. Available
abnormal state commands are listed below.
MS704:ABNormal:LIMits[:ALL]
This command will run all abnormal state operation mode tests.
Command Syntax MS704:ABNormal:LIMits[:ALL]
Parameters none
Examples MS704:ABN:LIM
Query Syntax none
MS704:ABNormal:LIMits:NOMinal
This command will run the abnormal state nominal operation mode test.
Command Syntax MS704:ABNormal:LIMits:NOMinal
Parameters none
Examples MS704:ABN:LIM:NOM
Query Syntax none
MS704:ABNormal:LIMits:LOW
This command will run the abnormal state low operation mode test.
Command Syntax MS704:ABNormal:LIMits:LOW
Parameters none
Examples MS704:ABN:LIM:LOW
Query Syntax none
MS704:ABNormal:LIMits:HIGH
This command will run the abnormal state high operation mode test.
Command Syntax MS704:ABNormal:LIMits:HIGH
Parameters none
Examples MS704:ABN:LIM:HIGH
Query Syntax none
MS704:ABNormal:VOLTage:TRANsient[:ALL]
This command will run all the abnormal voltage transients tests.
Command Syntax MS704:ABNormal:VOLTage:TRANsient[:ALL]
Parameters none
Examples MS704:ABN:VOLT:TRAN
Query Syntax none
Vista de página 215