Load Module Remote Programming ReFlex Power™ Programming Manual
M380056-03 Rev L 117
:PROTection <NRf>
Set the initial undervoltage protection value
:PROTection?
Return the initial undervoltage protection value
:VOLTage
:PROTect?
Get the initial over voltage protection value
:PROTect <NRf>
Set the initial over voltage protection value
:INPut
Calibrate Input subsystem
:CURRent
Calibrate Input Current subsystem
:COUNTS <NRi>
Enter the input current counts
:CALCulate
Calculate the current calibration
:FIVEPOINT <1-5> <NRf>
Set the input current value for each calibration point
:
RANGE <bool>
Change the current range
:CURRent <1-2> <NRf>
Set the input current calibration points
:CONDuctance
Calibrate Conductance subsystem
:COUNTS <NRi>
Set the conductance counts
:CALCulate
Calculate the conductance calibration
:VOLTage <1-2> <NRf>
Set the input voltage calibration points
:MEASure
Calibrate Measure subsystem
:OFFSET
Calibrate Measure Offset subsystem
:CALCulate
Calculate the voltage calibration
:UNLock <string>
Unlock the calibration data file; enables storage of the calibration
constants. The enable string is <“6867”>.
:STORe
Stores the calibration constants.
:LOCK
Locks the calibration data file; disables calibration data store, thereby
preventing storage of the calibration constants.
5.7
MEASURE SCPI COMMANDS
5.7.1 MEASURE
SCPI
C
OMMANDS
S
UMMARY
MEASure<n>
:CURRent?
:VOLTage?
:
POWer?
5.7.2 MEASURE
C
OMMANDS
R
EFERENCE
MEASure<n>
MEASURE subsystem of commands
:CURRent?
Returns the floating-point value of the DC input current.
:VOLTage?
Returns the floating-point value of the DC input voltage.
:
POWer?
Returns floating point value of DC power